Customer Service Center
Home
Knowledge Base
Contact Support
License Activation
|
Contact Support
|
License Activation
|
Spirent KB Article
Doc ID: FAQ14982
Printer Friendly
Email Article Link
GNSS Positioning Simulator: Is there a risk that the mains input voltage can appear at the RF output port?
Environment/Versions
GSS8000
Answer
No there is no risk of mains supply appearing at the RF output.
Any mains fault would be isolated to the main supply and rupture the main input fuse.
Additional information
The power supply unit is isolated from the internal electronic which uses low voltage DC.
The RF path is insulated from the main electronics and also the main chassis.
Version
4.0
Publish Date
2016-02-12
Categories
Product : General Navigation,Positioning
Related Articles