VisionWorks Probe: Why is test attempt failing with "CIR EXCEEDS LIMIT. RETRY LATER."?
Environment/Versions
VisionWorks Probe - all versions
Answer
This indicates that your test will cause the maximum bandwidth available by the Probe to be exceeded. If this is a new install into an existing chassis, the probe may have the incorrect TOTALCIR and TOTALKFPS settings. To check and correct setting:
Connect to 7010 port of probe.
Issue: "RTRV-ATTR-EQPT:::CTAG;"
Note the TOTALCIR and TOTALKFPS settings.
For 7531 settings are: "TOTALCIR=1000 TOTALKFPS=1489".
For 7521 and 7522 not optioned for 20G settings are: "TOTALCIR=10000 TOTALKFPS=14881".
For 7522 optioned for 20G settings are: "TOTALCIR=20000 TOTALKFPS=29762"
For 7561 settings are: "TOTALCIR=100000 TOTALKFPS=148810".
If you find the settings are incorrect, correct by:
Stop all tests to probe including ADDR-LPBK and TWAMP-SERVER.
Connect to 7010 port of probe.
Issue: ACT-USER::SUPER:CTAG::SUPER;
Issue: ED-ATTR-EQPT:::CTAG::TOTALCIR=xxxx,TOTALKFPS=yyyy;
Where xxxx is the TOTALCIR and yyyy is the TOTALKFPS for the type of probe.
Issue: INIT-SYS:::CTAG::4;
Done.
If the setting is correct, the problem is that the addition of your test will exceed the totals above.
FAQ18657
has information on checking the CIR and KFPS current usage.