why my system diagnostics can not pass for step 28?

Answer

When you do system calibration for TASKIT/C2K or TestDrive-EVDO, you may encounter some problem for system diagnostics parts, the typical error is out of range error as below picture shows:

This error is mostly because you select the wrong AirAccess mode in the instrument config window, for the correct setting, you should select from AirAccess:

Instrument->Enter configuration->AirAccess I/O->Test Configuration Unit, the screen capture should be as below:


Product : TestDrive EV-DO,TASKIT C2K,TestDrive EV-DO,TASKIT C2K