Printer FriendlyEmail Article Link

How to make configure correlation coefficent to be 1 for multiple signal in SR5500 TestKit?

Environment/Versions

SR5500 TestKit

Answer

If you want to make the correlation coefficient to be 1 for multiple signals, you need the following procedure

1, Open SR5500 TestKi, from configuration - System /communicaiton setup.

(1) goto System - System Mode: Number of SR5500 instruments controlled by TestKit:  select the numbers of the SR5500 you want to control.

(2) goto Synchronization mode : Synchronize Play/Pause/Stop across units: Active

(3) Correlation coefficient : System based, Correlection Coefficent type:  Complex

2, Click ok and after several seconds, the settings are applied, goto Correlation Editor, Correlation Mode : Complex

You will need to input the coefficient for every line to be 1.0000+0.0000j, apply after the value have been changed.

 

 

 


Product : SR5500,SR5500